The NanoPhotometer UV/Vis spectrophotometer
Complete solution for
ultra low and standard volume applications |
|
The NanoPhotometer is
designed to provide the utmost flexibility in photometry.
The outstanding technology combines ultra low and standard volume applications for
improved concentration measurement and estimation of purity in one UV/Vis
spectrophotometer. The fully developed software enables more user comfort with the
established methods and assists with the predefined Functions self-programming to meet
future needs. Various photometric functions, e.g. single and multi wavelength measurement
combined with kinetic methods, opens up a wide range applications. The full wavelength
scan (190 nm - 1100 nm) allows curve interpretation for attainment of more detailed
scientific knowledge over whole spectrum. |
 |
| Advanced Technology |
|
The NanoPhotometer spectrophotometer is a simple-to-use
UV/Visible instrument. It is the perfect composition in highest quality to cover all
photometric applications in molecular biology, biochemistry, and microbiology.
The CCD array detector (1024 pixels) and the pulsed xenon lamp guarantee most accurate
measurements. The spectrophotometer has no moving parts, which is the basis of the rapid
scanning operating system within the range from 190 nm and 1100 nm.
The high performance instrument operates with wavelength reproducibility (± 1 nm) and
accuracy (± 2 nm) at a spectral bandwidth of 5 nm. Within the photometric range of -
0.3 to 2.5Absorbance, respectively 0 to 199%T, the NanoPhotometerTM performs with highest
linearity and reproducibility.
The back-lit liquid crystal display is very easy to navigate with the alphanumeric entry
and navigation arrow keys on the hard wearing, spill proof membrane keypad
After switching on the instrument will perform a series of self-diagnostic checks for
wavelength calibrations.
|
 |
Planview of the
NanoPhotometer Function
The sample is pipetted onto the centre of the measuring window. Due to the integrated
beam deflexion and the use of fibre-optic light conductors the sample can be measured
directly on the measurement window. Using one of the four available lids of the microliter
cell (lid 10, lid 50 and optional lid 5 or lid 100) different liquid columns of defined
pathlength are created. Virtual dilutions of up to 1/100 are generated in comparison to a
standard 1 cm cuvette measurement, without a manual dilution of the sample. Dilution
errors are avoided and the sample can be retrieved after the measurement. |

|
Sample analysis made easy |
|
LabelGuard Applications
for ultra low sample volume measurementsCuevette
Applications
for standard sample volume measurements
Functions
for self-programming methods
User Methods
for storage ofself-configured methods
Utilities
Instrument set up
|
 |
Applications |
Nucleic acids:
dsDNA, ssDNA, RNA, and OligonucleotidesLabeling:
efficiency dye incorporation rate for microarray experiments
Proteins:
Bradford, Lowry, BCA, Biuret, and UV Protein (A280)
Cell density:
for microbiology and cell culture

For more user comfort, only three steps are necessary to enter method of choice.
|
| The NanoPhotometer is equipped with pre-programmed
Applications for highest user comfort.
| Nucleic acids |
dsDNA, ssDNA, RNA, and Oligonucleotides, Labeling efficiency |
| Proteins |
Bradford, Lowry, BCA, Biuret, and UV Protein (A280) |
| Cell density |
for bacterial cell cultures |
Nucleic Acids Application

Nucleic Acids Quantification
It is well established that for determination nucleic acid concentration in solution
the absorbance at wavelength 260 nm (A260) is used. The function describing the
concentration to absorbance relation is the Lambert-Beer Law: A = e * c * d.
The absorbance (A) is the product of the substance specific extinction coefficient (e),
the concentration of the absorbing sample (c), and the optical pathlength in cm
(d).
A solution of dsDNA in a 10 mm pathlength cell with an optical density of 1.0 has a
concentration of 50 µg/ml.
The NanoPhotometer uses the factors 50, 40, 37 and 33 as defaults for dsDNA,
ssDNA, RNA and oligonucleotides, respectively, and compensates factors for dilution and
varying pathlength .
Nucleic Acids Estimation of Purity
Depending on the extraction/purification or synthesis/purification method of the
nucleic acids different impurities can be expected (TRIzol, humic acids, carbohydrates,
Guanidine thiocyanate, nucleotides, peptides, EDTA, phenol and protein). It is recommended
to include OD ratio measurement (A260/A280 and A260/A230) for purity estimation.
Ratio A260/A280
In solution, pure DNA and RNA typically have A260/A280 ratios of 1.8 and 2.0. If the
absorbance ratio is significantly less, the nucleic acid is probably contaminated with
protein. Accurate quantification of nucleic acid is not reliable without prior
purification, and the efficacy of this can be judged by the A260 /A280 ratio.
Ratio A260/A230
For RNA samples the ratio values <2.0 point out genomic DNA contamination.
Successful DNase I treatment displays in ratio values > 2.0.
Ratio values <1.5 indicate impurities of extraction chemicals or incompletely
removed constituents of cells.
Use of Background Correction
Background correction at a wavelength totally separate from the nucleic acid and
protein peaks at 260 and 280nm, respectively, is sometimes used to compensate for the
effects of background absorbance. The wavelength used is 320 nm and it can allow for the
effects of turbidity, high absorbance buffer solution and the use of reduced aperture
cells.
The NanoPhotometer offers the option of background correction at 320 nm for
nucleic acid determination.
Dye incorporation rate
To determine the dye incorporation rate, the absorbance reading at the wavelength
reported for maximum absorbance of the fluorescence dye is used. The corresponding
extinction coefficient of the dye is used in the Lambert-Beer Law to determine the dye
concentration (c = A / (e * d)). Comparing these values with the DNA concentration gives a
dye incorporation rate.
FOI = C(dye) / C(nucleic acid)
Example: Frequency of Incorporation (FOI) of Cy3 per 1000 bases:
FOI(Cy3) = 58.5 * A550/A260
The user can switch between two options of result presentation: a data table or a scan
plot.
Use of Background Correction
The absorbance reading of dye labelled nucleic acids at 260 nm is affected by the dye
contribution. To obtain accurate concentration values, the contribution has to be
eliminated using correction factors.
The NanoPhotometer offers the option to select the background correction for
dye contribution of absorbance reading at 260 nm for nucleic acid determination.
Protein Applications
1. Protein UV
| The Protein UV method determines Proteins at 280nm. The
effect of nucleic acid in the protein solution due to strong nucleotide absorbance at 280
nm can be compensated by measuring Abs 260, and following the equation of Christian
Warburg:
Protein (mg/ml) = 1.55*A280 - 0.76* A260
The use of background correction at 320 nm is optional. |
 |
2. BCA
| The BCA method depends on reaction between Cupric ions and
peptide bonds, but in addition combines this reaction with the detection of Cuprous ions
using bicinchoninic acid (BCA), giving an absorbance maximum at 562 nm. The BCA process
is less sensitive to the presence of detergents used to break down cell walls.
Select units of measurement:
µg/µl, pmol/µl, µg/ml, mg/dl, µg/l, mg/l, g/l, mmol/l,
µmol/l, U/l, %, ppm, ppb, conc, none.
Calibration mode:
Either measurements of prepared standards or manually enter data via keypad.
The Calibration Screen shows the calibration values and allows standards to be
measured. A graph will display the results and the fitted curve as measurements are made.
Poor reading can be repeated (with replicates on). |




 |
The user can switch between two options of result presentation: a data table or a scan
plot.
3. Bradford
| The Bradford method depends on quantitating the binding of
a dye, Coomassie Brilliant Blue, to an unknown protein and comparing this binding to that
of different, known concentrations of a standard protein at 595 nm (default setting); this
is usually BSA, bovine serum albumin. Select units of measurement:
µg/µl, pmol/µl, µg/ml, mg/dl, µg/l, mg/l, g/l, mmol/l,
µmol/l, U/l, %, ppm, ppb, conc, none.
Calibration mode:
Either measurements of prepared standards or manually enter data via keypad
The Calibration Screen shows the calibration values and allows standards to be
measured. A graph will display the results and the fitted curve as measurements are made.
Poor reading can be repeated (with replicates on).
|






|
The user can switch between two options of result presentation: a data table or a scan
plot.
4. Lowry
| The Lowry method depends on quantifying the colour obtained
from the reaction of Folin-Ciocalteu phenol reagent with the tylsryl residues of an
unknown protein and comparing with those derived from a standard curve of a standard
protein at 750nm; this is usually BSA, bovine serum albumin Select units of measurement:
µg/µl, pmol/µl, µg/ml, mg/dl, µg/l, mg/l, g/l, mmol/l,
µmol/l, U/l, %, ppm, ppb, conc, none.
Calibration mode:
Either measurements of prepared standards or manually enter data via keypad
The Calibration Screen shows the calibration values and allows standards to be
measured. A graph will display the results and the fitted curve as measurements are made.
Poor reading can be repeated (with replicates on).
|





 |
The user can switch between two options of result presentation: a data table or a scan
plot.
5. Biuret
| The Biuret method measures the reaction between Cupric ions
and peptide bonds in an alkali solution, resulting in the formation of a complex absorbing
at 546 nm. Select units of measurement:
µg/µl, pmol/µl, µg/ml, mg/dl, µg/l, mg/l, g/l, mmol/l,
µmol/l, U/l, %, ppm, ppb, conc, none.
Calibration mode:
Either measurements of prepared standards or manually enter data via keypad
The Calibration Screen shows the calibration values and allows standards to be
measured. A graph will display the results and the fitted curve as measurements are made.
Poor reading can be repeated (with replicates on). |





 |
The user can switch between two options of result presentation: a data table or a scan
plot.
OD 600 Cell Density Applications
| The method OD 600 is used to determinate the absorbance at
600 nm. Bacterial cell cultures are routinely grown until the absorbance at 600 nm (known
as OD600; default setting) reaches approximately 0.4 prior to induction or harvesting. A
linear relationship exists between cell number (density) and OD 600 up to approx. 0.6. Select
units of measurement:
OD, cells/ml.
With units OD selected the results output are OD values. If cells/ml was selected two
further parameters are displayed (Factor and Multiplier).
The measured absorbance in these turbid samples is due to light scattering, and not
the result of molecular absorption.
A calibration curve can be determined by comparing measured OD 600 to expected OD 600
determined by using an alternative technique (e.g. microscope slide method). |
 
|
|
|
Open for future development |
|
| For special photometric applications or future development, a
selection of pre-defined features for general photometric functions enables to create own
methods. The configured methods can be saved in nine empty folders (nine methods per
folder) under User Methods. The method folder names are editable to allow
self-organisation. |
 |
Complete solution for ultra low and standard volume applications
| The NanoPhotometerTM is equipped with a selection
of pre-defined general photometric features to provide the utmost flexibility in
photometry. Scientists are enabled to configure own user methods easily. |
 |
1. Single Wavelength
| The feature Single Wavelength allows simple
absorbance (Abs) or transmission (%T) measurements at a single user defined wavelength.
The result windows shows the amount of light passed through a sample relative to a
reference. |
 |
2. Concentration
The feature Concentration
determines the concentration of samples from a single wavelength reading as a
Colorimetric assay.
The amount of light that has passed through a sample
relative to a reference is multiplied by a factor to obtain the concentration. The factor
may be a known figure, or may be calculated by the instrument by measuring a standard of
known concentration.
The units to present the results are selectable (µg/ml, µg/µl, pmol/µl, mg/dl,
mmol/l, µmol/l, g/l, mg/l, µg/l, U/l, %, ppm, ppb, conc or none). |
 |
3. Wavescan
The feature Wavescan records
an absorption (Abs) or transmission (%T) spectrum between two user defined wavelengths,
maximum range 190 nm to 1100nm.
The spectral plot allows simple identification of peak
height and position.
Benefit
The curve interpretation of the entire spectrum provides
most detailed information for a sample. |
 |
4. Kinetics
The feature Kinetics performs
a colorimetric assay which follows the change in absorbance as a function of time at fixed
wavelength.
Kinetic studies are routinely used for the enzymatic
determination of compounds in food, beverage and clinical laboratories by measuring
NAD/NADH conversion at 340nm.
The change of absorbance per minute (A/min),
concentration (A/min x factor) and correlation coefficient (calculated from
a best fit of the data points) are displayed. |
 |
5. Standard Curve
The feature Standard
Curve is a colorimetric assay at a single wavelength based on a user programmed
curve.
The preparation of a multi-point calibration curve from
standards of known concentration to quantify unknown samples is a fundamental use of a
spectrophotometer; this instrument has the advantage of being able to store this curve as
a method, using up to 9 standards.
Besides, the data value used for the Standard Curve
Calibration can be entered manually.
The units to present the results are selectable: µg/ml,
µg/µl, pmol/µl, mg/dl, mmol/l, µmol/l, g/l, mg/l, µg/l, U/l, %, ppm, ppb, conc or
none. |
 
|
6. Multi Wavelength
| The feature Multi Wavelength determines
absorbance (Abs) or transmission readings at up to 5 user defined wavelengths relative to
a reference. The results are presented in a scan plot covering the range of wavelengths
selected and highlighted the relevant wavelength with cursors and a table of values. |
 |
7. Absorbance Ratio
| The feature Absorbance Ratio calculates the
ratio of absorbance readings from two user defined wavelength. |
 |
The amount of light that has passed
through a sample relative to a reference at two wavelengths. Volume of the sample and
dilution factor can be entered. The unit to present results are selectable: µg/ml,
ng/µl, µg/µl.
The NanoPhotometer offers the possibility for user-configured methods.
Technical Specifications
Photometric
specifcations |
Photometric mode |
Absorbance, %Transmission, concentration, scan, ratio, multi
and single wavelength, kinetics in ?Abs*factor/min |
Kinetics mode |
Start time, end time, interval, numbers of points TBA. |
Analysis mode |
LSQ, rate, factor, curve |
Scan range |
Full wavelength scan 200nm 950nm |
|
Wavelenght range 190nm - 1100nm |
| Slew rate |
Measurement time for full spectrum acquired simultaneously in
less then <5 sec |
| Photometric range |
-0.3 2.499 Abs0-199% T |
| Photometric Reproducibility |
±0.003Abs (0 to 0.5Abs)±0.007Abs (0.5-1.0Abs) @ 260nm |
| Photometric Accuracy |
±0.005A or ±1% of the reading, whichever is the greater |
| Wavelength reproducibility |
< ±0.2nm |
| Wavelength accuracy |
± 2 nm |
| Zero stability |
±0.003Abs/hour after 20min warm up @ 340nm |
| Noise |
0.002A rms at 0A @ 260nm0.005A (pk to pk) at 0A @ 260nm |
| Bandwidth |
5 nm |
| Stray Light |
<0.5% at 220nm and 340nm NaNO2 |
|
|
|
|
Optical
arrangements |
| Optical arrangements |
Dual channel Czerny Turner with flat grating, 1024 pixel CCD
array |
| Lamp |
Xenon flash lamp |
| Lifetime |
109 flashes |
| Warranty |
1 year |
| Performance Verification |
Auto diagnostics when switched on |
|
|
|
|
Methods
Programming |
| Method storage |
Up to 90 user-defined methods |
| Built-in methods for |
nucleic acids (dsDNA, ssDNA, RNA, Oligos),
labelling efficiency (dye incorporation for microarrays),
proteins (Bradford, Lowry, BCA, Biuret, UV),
bacterial cell density (microbiology) |
| Predefined functions for |
Self-programming methods:
single and multiple wavelength, concentration, wavelength scan,
kinetics, standard curve, and ratio calculation |
|
|
|
|
Standard
Cuvettes |
| Cuvettes cells |
10mm cuvette cells with outside dimension 12.5mm x 12.5mm.
Quartz SUPRASIL cells are recommended. |
| Cuvette compartment |
10mm cell (outside dimension 12.5mm x 12.5mm) |
| Cuvette storage |
8 x 10mm cells |
|
|
|
|
Electrical |
| Power |
External autoranging power supply 225 VA |
| Operating voltages |
90 - 250 V, (50 Hz / 60 Hz), Max 30VA |
|
|
|
|
Operating
environment |
| Temperature |
10 - 35 °C |
| Cooling |
Natural air, no internal fan |
| Heated cell holder |
Not necessary |
|
|
|
|
Input /
Output Ports |
| USB |
USB is
standard for connection to a PVC for direct data download for spreadsheet calculations,
printout and data storage |
| Built-In Printer |
Internal
Serial interface to optional external printer.
Graphics to be printed out on thermal high stability paper, 56mm wide, and 40 columns |
| BlueTooth |
Internal Serial interface to optional Blue Tooth wireless
link for connection to a PC |
Safety
The instruments is designed to conform with:
EN-61010-1:2001 and EN 6010-2-101: 2002
Certification
To Article 5 of the Low Voltage Directive (73/23/EEC), the EMC Directive (89/336/EEC)
and the IVD Directive
The standards used shall be
| EN 61010-1: 2001 |
safety requirements for electrical equipment for measurement,
control and laboratory use |
| EN 61326-2.3: 1998 |
electromagnetic compatibility - generic emission standard
Electrical equipment for measurement, control and laboratory use |
| EN 61000-4-6: 1992 |
electromagnetic compatibility - generic immunity standard
part 1. Residential, commercial and light industry |
|